Iranian Journal of Radiation Research. 2005; 3 (2): 73-78
in English | IMEMR | ID: emr-71088

ABSTRACT

A prototype stereotactic radiosurgery set was designed and constructed for a Neptun 10 PC linac that is currently being used at Imam Reza hospital in Mashhad. A complete quality assurance program was designed and performed for the constructed system including isocentric accuracy test, localization accuracy test, dose delivery accuracy test and leakage radiation test. Target simulator, control alignment device and plexiglass phantom which were parts of the developed hardware were used to fulfill quality assurance program. The average isocentric shift resulted from the gantry rotation and couch turning were respectively obtained to be 1.4 and 2 mm. The average localization error in the three coordinates was found to be 2.2 mm. The total treatment uncertainty due to all of the probable errors in the system was equal to 4.32 mm. The dose delivery accuracy test was carried out, the result indicated a 3-7% difference between the given and measured dose. The quality assurance tests showed consistent performance of the constructed system within the accepted limits; however, some inconsistency might exist in certain cases. The safety of SRS method is increased when the overall uncertainty is minimized and the treatment of the lesions adjacent to critical organs is avoided


Subject(s)
Radiosurgery/standards , Quality Assurance, Health Care/methods , Quality Assurance, Health Care/statistics & numerical data , Radiometry