Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Journal of Pharmaceutical Analysis ; (6): 190-199, 2015.
Article in Chinese | WPRIM | ID: wpr-467109

ABSTRACT

An analytical method based on solid phase extraction was developed and validated for analysis of adefovir in human plasma. Adefovir-d4 was used as an internal standard and Synergi MAX RP80A (150 mm × 4.6 mm, 4μm) column provided the desired chromatographic separation of compounds followed by detection with mass spectrometry. The method used simple isocratic chromato-graphic condition and mass spectrometric detection in the positive ionization mode. The calibration curves were linear over the range of 0.50–42.47 ng/mL with the lower limit of quantitation validated at 0.50 ng/mL. Matrix effect was assessed by post-column infusion experiment to monitor phospholipids and post-extraction addition experiment was performed. The degree of matrix effect for adefovir was determined as 7.5%and ion-enhancement in five different lots of human plasma was 7.1%and had no impact on study samples analysis with 4.5 min run time. The intra- and inter-day precision values were within 7.7% and 7.8%, respectively, for adefovir at the lower limit of quantification level. Validated bioanalytical method was successfully applied to clinical sample analysis.