Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Chinese Journal of Tissue Engineering Research ; (53): 3445-3450, 2014.
Article in Chinese | WPRIM | ID: wpr-446588

ABSTRACT

BACKGROUND:Good position of acetabular prosthesis during total hip replacement plays a key role in success of the operation. Traditional location tool has poor accuracy, is too complicated, and limits its application. OBJECTIVE:To compare the advantages and disadvantages of gyroscope with the traditional technique in acetabular component orientation in total hip arthroplasty. METHODS:Using lateral approach, physicians with different experiences used traditional technique to locate acetabular cup 100 times (20 times in each person; 40° abduction and 15° anteversion), and then used gyroscope to locate 60 times at the same angle in the same plastic hip model. The error between the planed and measured values was recorded. RESULTS AND CONCLUSION: Compared to the traditional technique, the error of the abduction and anteversion decreased obviously. These indicated that the gyroscope can make the orientation of acetabular component more precise in total hip arthroplasty.

SELECTION OF CITATIONS
SEARCH DETAIL