1.
Philippine Journal of Ophthalmology ; : 66-72, 2004.
Article in English | WPRIM | ID: wpr-632335

ABSTRACT

Methods: Patients with and without glaucoma underwent a complete eye evaluation, automated perimetry, scanning laser polarimetry with the GDx 400, and opticdisc photography. Two glaucoma experts graded each study eye. Two-by-two tables were constructed for 5 GDx parameters (average thickness, superior average, inferior average, ellipse average, and ellipse modulation) and the CDx number. Receiver operating characteristic (ROC) curves were generated. Results: The study included 355 patients (171 normal, 184 glaucoma). The mean values of the 5 GDx parameters were lower for the glaucoma than for the normal group. The sensitivity and specificity of the GDx 400 were 45.4 percent and 91.9 percent if the cut off level of the GDx number was 71. Ellipse modulation (EM) measures have the best ROC curve with area under the curve of 0.725. Conclusion: The GDx 400 nerve-fiber analyzer is primarily used as a screening tool to detect the presence or absence of glaucoma. Its accuracy can be improved with use of continuous corneal compensator.


Subject(s)
Humans , Male , Female , Aged , Middle Aged , Adult , Glaucoma , Retina , Nerve Fibers , Scanning Laser Polarimetry