Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
1.
Annals of the Academy of Medicine, Singapore ; : 388-393, 2007.
Article in English | WPRIM | ID: wpr-250810

ABSTRACT

<p><b>INTRODUCTION</b>The new 3 Tesla (T) magnetic resonance (MR) scanners yield improved signal-to-noise ratio and spatial resolution with superior background suppression compared to lower field strength systems. This is advantageous for MR angiograms. The purpose of our study was to compare unenhanced three-dimensional time-of-flight magnetic resonance angiography (3D TOF MRA) at 3T with catheter digital subtraction angiography (DSA) in detecting unruptured intracranial aneurysms.</p><p><b>MATERIALS AND METHODS</b>Out of 1375 consecutive patients who underwent unenhanced 3D TOF MRA at 3T, 15 patients with unruptured intracranial aneurysms were retrospectively identified. Nine of these 15 patients had DSA as the reference standard for comparison. Aneurysm size, location and morphology were independently assessed on both MRA and DSA by 2 radiologists.</p><p><b>RESULTS</b>Seventeen aneurysms ranging in size from 1 mm to 24 mm were identified in 15 patients on MRA. DSA confirmed the aneurysms in 9 patients with good anatomical correlation compared with the MRA findings.</p><p><b>CONCLUSIONS</b>3D TOF MRA at 3T has good correlation with DSA and aneurysms as small as 1 mm in size can be detected. This can be a promising, non-invasive method for aneurysm surveillance.</p>


Subject(s)
Adult , Aged , Female , Humans , Male , Middle Aged , Angiography, Digital Subtraction , Image Processing, Computer-Assisted , Methods , Intracranial Aneurysm , Diagnosis , Magnetic Resonance Angiography , Methods , Retrospective Studies , Singapore
SELECTION OF CITATIONS
SEARCH DETAIL