Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add filters








Year range
1.
The Journal of Advanced Prosthodontics ; : 183-190, 2015.
Article in English | WPRIM | ID: wpr-194331

ABSTRACT

PURPOSE: This study evaluated the fracture load of customized zirconia abutments with titanium insert according to preparation depths, with or without 5-year artificial aging. MATERIALS AND METHODS: Thirty-six identical lithium disilicate crowns (IPS e.max press) were fabricated to replace a maxillary right central incisor and cemented to the customized zirconia abutment with titanium insert on a 4.5x10 mm titanium fixture. Abutments were fabricated with 3 preparation depths (0.5 mm, 0.7 mm, and 0.9 mm). Half of the samples were then processed using thermocycling (temperature: 5-55degrees C, dwelling time: 120s) and chewing simulation (1,200,000 cycles, 49 N load). All specimens were classified into 6 groups depending on the preparation depth and artificial aging (non-artificial aging groups: N5, N7, N9; artificial aging groups: A5, A7, A9). Static load was applied at 135 degrees to the implant axis in a universal testing machine. Statistical analyses of the results were performed using 1-way ANOVA, 2-way ANOVA, independent t-test and multiple linear regression. RESULTS: The fracture loads were 539.28 +/- 63.11 N (N5), 406.56 +/- 28.94 N (N7), 366.66 +/- 30.19 N (N9), 392.61 +/- 50.57 N (A5), 317.94 +/- 30.05 N (A7), and 292.74 +/- 37.15 N (A9). The fracture load of group N5 was significantly higher than those of group N7 and N9 (P<.017). Consequently, the fracture load of group A5 was also significantly higher than those of group A7 and A9 (P<.05). After artificial aging, the fracture load was significantly decreased in all groups with various preparation depths (P<.05). CONCLUSION: The fracture load of a single anterior implant restored with lithium disilicate crown on zirconia abutment with titanium insert differed depending on the preparation depths. After 5-year artificial aging, the fracture loads of all preparation groups decreased significantly.


Subject(s)
Aging , Axis, Cervical Vertebra , Crowns , Dental Implants , Incisor , Linear Models , Lithium , Mastication , Titanium
2.
The Journal of Korean Academy of Prosthodontics ; : 331-337, 2014.
Article in Korean | WPRIM | ID: wpr-201567

ABSTRACT

In conventional bar- and clip-retained overdentures, all errors during impression making and cast fabrication result in non-passive fit of bars. SFI bar is prefabricated and assembled in the patient's mouth without the use of soldering, laser welding or conventional bonding techniques, thus reducing stress transmission to, bone loss around the implants and leading to fewer errors and lower costs. A clinical case will be presented below to demonstrate the use of the SFI Bar (Stress Free on Implant Bar) to restore an severe atrophy edentulous maxilla. In this case, no lateral stress could be applied to the implants due to the telescopic design of the bar joints. However, periodic recall check is necessary and long-term clinical results are required.


Subject(s)
Atrophy , Denture, Overlay , Joints , Maxilla , Mouth , Mouth Rehabilitation , Welding
SELECTION OF CITATIONS