Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Chinese Journal of Medical Instrumentation ; (6): 292-295, 2022.
Article in Chinese | WPRIM | ID: wpr-928907

ABSTRACT

The treatment of refractory Glaucoma is a difficult problem in clinical ophthalmology. For refractory glaucoma patients with hyphema, shallow anterior chamber, anterior conglutination of peripheral chamber angle, corneal endothelium dystrophy or decompensated, at present, there is no effective treatment. In order to solve this problem, a new type posterior integral glaucoma valve with IOP control device was designed using medical titanium alloy, and the valve model was established by Abaqus software, and the stiffness and preload of the valve were analyzed by finite element method. The results showed that the opening and closing of the valve were controlled automatically by the pressure difference between the front and back of the valve, and the opening and flow rate of the valve increase dynamically with the increase of intraocular pressure, and finally reached the set ideal IOP value of steady state.


Subject(s)
Humans , Finite Element Analysis , Follow-Up Studies , Glaucoma , Glaucoma Drainage Implants , Intraocular Pressure , Treatment Outcome
SELECTION OF CITATIONS
SEARCH DETAIL