The Journal of the Korean Orthopaedic Association ; : 746-751, 2008.
Article in Korean | WPRIM | ID: wpr-646477

ABSTRACT

PURPOSE: The objective of this study was to compare the bonding strength between cement and an implant according to the cementing time. MATERIALS AND METHODS: The two types of cement used were CMW(R) 1 and 3 (Depuy Ltd., Blackpool, UK). Plastic molds containing CMW(R) 1 were pressed onto metal blocks at 2, 4, 5 or 6 minutes after mixing the cement, while molds containing CMW(R) 3 were pressed onto blocks at 3, 5, 6 or 7 minutes after mixing the cement. Tensile strength was tested with using an Instron Model 8874 (Instron Corp., Canton, MA, USA). Tensile strengths were compared using the Kruskal-Wallis test and the Mann-Whitney U test. RESULTS: The strongest bonding strengths for the tensile load were at 2 minutes post-mixing for CMW(R) 1, and at 3 minutes post-mixing for CMW(R) 3. The strength rapidly decreased after 5 minutes for CMW(R) 1 and after 6 minutes for CMW(R) 3. CONCLUSION: This study suggests that the risk of loosening between cement and an implant is likely to be minimized by the surgical technique that considers the bonding strength according to time.


Subject(s)
Arthroplasty , Fungi , Plastics , Tensile Strength