Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Chinese Journal of Radiology ; (12)2001.
Article in Chinese | WPRIM | ID: wpr-679639

ABSTRACT

Objective To evaluate the quality of the portal images acquired by computed radiography(CR)system and conventional screen-film system,respectively.Methods imaging plates (IP)and X-ray films of a home-devised lead phantom with a leakage of 6.45% were acquired,and modulation transfer function(MTF)curves of the both images were measured using edge method.Portal images of 40 nasopharyngeal cancer patients were acquired by IP and screen-film system respectively.Two doctors with similar experience evaluated the damage degree of petrosa] bone,the receiver operating characteristic(ROC)curve of CR images and general images were drawn according to two doctors evaluation results.Results The identification frequency of CR system and screen-film system were 1.159 and 0.806 Lp/mm respectively.For doctor one,the area under ROC curve of CR images and general images were 0.802 and 0.742 respectively.For doctor two,the area under ROC curve of CR images and general images were 0.751 and 0.600 respectively.The MTF curve and ROC curve of CR are both better than those of screen-film system.Conclusion The image quality of CR portal imaging is much better than that of screen-film system.The utility of CR in linear accelerator for portal imaging is promising in clinic.

SELECTION OF CITATIONS
SEARCH DETAIL