Journal of Neurogastroenterology and Motility ; : 455-462, 2020.
Article | WPRIM | ID: wpr-833891

ABSTRACT

Background/Aims@#Esophageal baseline impedance (BI) can be extracted from pH-impedance tracings as mean nocturnal baseline impedance (MNBI), and from high-resolution impedance manometry (HRIM), but it is unknown if values are similar between acquisition methods across HRIM manufacturers. We aim to assess correlations between MNBI and BI from HRIM (BI-HRIM) from 2 HRIM manufacturers in the setting of physiologic acid exposure time (AET). @*Methods@#HRIM and pH-impedance monitoring demonstrating physiologic AET ( 0.5 were seen at MNBI at 7 cm for both systems, and at 9 cm for Medtronic. DBI-HRIM correlated with MNBI at 3 cm and 5 cm (P 0.1). @*Conclusions@#While numeric differences exist between manufacturers, BI-HRIM correlates with MNBI from corresponding channels in patients with physiologic AET. Comparison with AET elevation is needed to determine correlations between pathologic MNBI with BI-HRIM across manufacturers. The optimal HRIM channels from which BI values should be extracted also warrants further study.