Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Article in English | IMSEAR | ID: sea-39527

ABSTRACT

OBJECTIVE: The aim of the present study was to evaluate a manual CD4+ count kit assay (CD4+: cytospheres) for CD4+ T-lymphocyte count compared with flow cytometric method in HIV infected patients. MATERIAL AND METHOD: One hundred thirty three HIV infected patients were recruited from the out patient department of Khon Kaen Hospital. Blood samples were done by a manual CD4+ count kit assay (CD4+: cytospheres) and flow cytometry for CD4+ T-lymphocyte count. The data were analyzed for diagnostic test and correlation coefficient. RESULTS: The data of cytospheres assay and flow cytomeric method showed good correlation (r = 0.88) for the total group. At the absolute CD4+ T-lymphocyte 200 cells/cu.mm, the cytospheres assay demonstrated sensitivity 83.10% (76.73-89.47%), specificity 93.55% (89.37-97.72%), PPV 93.65% (89.51-97.79%), NPV 82.86% (76.45-89.26%). In the case of CD4+ T-lymphocyte count were lower than 200 cells/cu.mm, the cytospheres assay displayed progressive decrease in sensitivity successive increase in specificity. CONCLUSION: The cytospheres technique is an alternative noncytofluorometric assay for CD4+ T-lymphocyte count. This test may be useful for screening in HIV infected adult patients in community hospitals where flow cytometry technique is not available. But the assay is limited in determination only absolute CD4+ T-lymphocyte count with higher than 30 cells/cu.mm. This technique is not benefit in pediatric HIV/AIDS patient due to percentage CD4+ value did not obtained. The quality control should be concern technical skill and proficiency testing for laboratory setting.


Subject(s)
CD4 Lymphocyte Count/methods , CD4-Positive T-Lymphocytes , Flow Cytometry , HIV Infections/diagnosis , Humans , Sensitivity and Specificity , Statistics as Topic
SELECTION OF CITATIONS
SEARCH DETAIL