Chinese Medical Equipment Journal ; (6): 63-67, 2018.
Article in Chinese | WPRIM | ID: wpr-700066

ABSTRACT

Object To verify and evaluate the performances of Sysmex XN-9000 hematology analyzer under body fluid mode when used for peritoneal and pleural fluids detection.Methods According to the guidance of ICSH, Sysmex XN-9000 hematology analyzer under body fluid mode had its performances verified on background counting, carryover, precision, linear range and etc of RBC-BF and TC-BF, and then was compared with microscopy on RBC-BF, TC-BF, PMN%, MN%, N%, L%, M%and E%.Results The RBC-BF and TC-BF had the background counting being 0.00, carryover being 0.00%and 0.07% respectively, coefficients of variation (CV) of within-run precision being from 0.96%to 17.89%as well as from 2.19%to 10.33%respectively, CV of between-run precision being from 3.34%to 4.73%as well as from 8.33%to 10.75%resptctively, and the linear ranges being (0~5) ×1012/L and (0~20) ×109/L respectively. There was a high correlation between Sysmex XN-9000 hematology analyzer and microscopy when detecting RBC-BF, TC-BF, PMN%, MN%, N%, L%, M%and E%. The Spearman rank correlation coefficients (rs) were 0.977, 0.995, 0.863, 0.929, 0.926, 0.949, 0.965 and 0.816 (P<0.05), and there were statistical significances between the correlations. Bland-Altman bias analysis on the analyzer and microscopy showed that RBC-BF had the bias (4.6%) lower than that (5.6%) of TC-BF;MN%had the bias (-2.0%) lower than that (4.2%) of PMN%;L%, N%, M%and E%had the bias being-0.5%, 4.4%,-5.9%and-1.6%respectively, which were all met the requirements of the manufacturer.Conclusion Sysmex XN-9000 hematology analyzer under body fluid mode proves its performances for routine detection of peritoneal and pleural fluids.

SELECTION OF CITATIONS
SEARCH DETAIL