Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Medical Principles and Practice. 2016; 25 (2): 159-162
in English | IMEMR | ID: emr-178539

ABSTRACT

Objective: The purpose of this study was to evaluate and compare the rate of bacterial contamination of reused and new unused burs after different sterilization sessions


Materials and Methods: The test group consisted of 40 used fissure burs, and the control group of 40 unused new fissure burs [total n = 80]. The burs from both groups were precleaned according to standard protocols and then subjected to two sterilization sessions [high- and low-steam pressure autoclaving]. After each sterilization session, the burs were transferred into incubation tubes which contained thioglycollate culture medium and were monitored daily for a period of 48 h to detect any bacterial growth. Data were collected and statistical analysis was done using Fisher's exact test


Results: Of the 40 burs of the test group, 2 burs [5%] showed positive bacterial growth, whereas no bur from the control group showed any sign of bacterial growth after high-pressure autoclaving. The colony structure and Gram staining were compatible with the growth of Staphylococcus epidermis . After a second sterilization session at low-pressure steam autoclaving, no bacterial growth was observed for the test group, but 1 bur [2.5%] from the control group showed bacterial growth and Gram-positive staining matched well with the growth of Brevibacterium species


Conclusions: The new and unused burs were 100% sterile after high-pressure steam autoclaving, whereas 5% of the reused burs appeared positive with bacterial contamination. After low-pressure steam autoclaving, reused burs were 100% sterile, but 1 new bur demonstrated bacterial contamination

SELECTION OF CITATIONS
SEARCH DETAIL