1.
JPMI-Journal of Postgraduate Medical Institute. 2015; 29 (4): 231-236
in English | IMEMR | ID: emr-179779

ABSTRACT

Objective: to compare bone marrow biopsy needle [BMBN] with bone biopsy needle [BBN] in CT guided core biopsies considering size of core, diagnostic yield and cost of needles


Methodology: retrospective study was done on 50 out of 68 patients referred to Interventional Radiology department for CT guided bone biopsy from 2008 till 2013. Grading of specimen length was done as small, medium and large. Diagnostic yield was determined on basis of clinical, imaging and histopathology data. Local cost of both needle types was determined. Comparison was made and statistical analysis was performed


Results: biopsy was performed by BBN in 16 patients and by BMBN in 34 patients. Overall diagnostic yield of bone biopsies was 82% [41 of 50 biopsies]. Diagnostic yield with BMBN was 85% and with BBN was 75%. Overall longer specimen length was obtained using BMBN. Statistically, there was no significant difference between them in diagnostic yield and specimen length; however, BMBN costs almost one half than BBN


Conclusion: diagnostic yield and sample length using bone marrow biopsy needle proved to be equivalent to bone biopsy needle. Bone marrow biopsy needle has an added benefit of being low in cost. Despite being a non-coaxial technique these needles can achieve diagnostic yield comparable to that of coaxial techniques as reported in past literature