Radiol. bras ; 48(6): 363-367, Nov.-Dec. 2015. tab, graf
Article in English | LILACS | ID: lil-771086

ABSTRACT

Abstract Objective: Derive filtered tungsten X-ray spectra used in digital mammography systems by means of Monte Carlo simulations. Materials and Methods: Filtered spectra for rhodium filter were obtained for tube potentials between 26 and 32 kV. The half-value layer (HVL) of simulated filtered spectra were compared with those obtained experimentally with a solid state detector Unfors model 8202031-H Xi R/F & MAM Detector Platinum and 8201023-C Xi Base unit Platinum Plus w mAs in a Hologic Selenia Dimensions system using a direct radiography mode. Results: Calculated HVL values showed good agreement as compared with those obtained experimentally. The greatest relative difference between the Monte Carlo calculated HVL values and experimental HVL values was 4%. Conclusion: The results show that the filtered tungsten anode X-ray spectra and the EGSnrc Monte Carlo code can be used for mean glandular dose determination in mammography.


Resumo Objetivo: Derivar espectros filtrados de raios X de tungstênio utilizados em sistemas de mamografia digital por meio de simulações Monte Carlo. Materiais e Métodos: Espectros filtrados por filtro de ródio foram obtidos para potenciais do tubo entre 26 e 32 kV. Os valores de camada semirredutora (CSR) dos espectros filtrados simulados foram comparados aos valores obtidos experimentalmente com um detector de estado sólido Unfors modelo 8202031-H Xi R/F & MAM Detector Platinum e unidade base 8201023-C Xi Platinum Plus w mAs em um sistema Hologic Selenia Dimensions utilizado no modo radiografia direta. Resultados: Os valores de CSR calculados mostraram boa concordância quando comparados com os valores obtidos experimentalmente. A maior diferença relativa entre os valores de CSR calculados com Monte Carlo e os valores de CSR experimentais foi 4%. Conclusão: Os resultados demonstram que os espectros filtrados de raios X de ânodo de tungstênio e o código de Monte Carlo EGSnrc podem ser utilizados para determinar a dose glandular média em mamografia.

2.
Radiol. bras ; 47(6): 361-367, Nov-Dec/2014. tab, graf
Article in English, Portuguese | LILACS-Express | LILACS | ID: lil-732753

ABSTRACT

Objetivo: Avaliar a dose glandular média utilizando um detector de estado sólido e a qualidade da imagem de um sistema de radiografia direta a partir de objetos simuladores. Materiais e Métodos: Irradiações foram realizadas utilizando o controle automático de exposição e placas de polimetilmetacrilato com diferentes espessuras para o cálculo da dose glandular média. A qualidade da imagem foi avaliada por meio das estruturas visualizadas nas imagens dos objetos simuladores. Resultados: Considerando a incerteza das medições, os resultados de dose glandular média estão de acordo com os valores fornecidos pelo equipamento e com os níveis de referência adotados internacionalmente. Os resultados obtidos a partir das imagens dos objetos simuladores estavam em conformidade com os valores de referência. Conclusão: Este trabalho contribui para verificar a conformidade do equipamento em relação a dose e qualidade da imagem. .


Objective: To evaluate the mean glandular dose with a solid state detector and the image quality in a direct radiography system, utilizing phantoms. Materials and Methods: Irradiations were performed with automatic exposure control and polymethyl methacrylate slabs with different thicknesses to calculate glandular dose values. The image quality was evaluated by means of the structures visualized on the images of the phantoms. Results: Considering the uncertainty of the measurements, the mean glandular dose results are in agreement with the values provided by the equipment and with internationally adopted reference levels. Results obtained from images of the phantoms were in agreement with the reference values. Conclusion: The present study contributes to verify the equipment conformity as regards dose values and image quality. .

3.
Radiol. bras ; 41(2): 109-114, mar.-abr. 2008. graf
Article in Portuguese | LILACS | ID: lil-482997

ABSTRACT

OBJETIVO: Esta pesquisa visou a avaliar a qualidade dos serviços de mamografia do Distrito Federal e o impacto de uma intervenção de inspeção e capacitação. MATERIAIS E MÉTODOS: Foram estudados 41 serviços de mamografia no Distrito Federal no período de 2000 a 2002. A intervenção consistiu na inspeção inicial seguida de um treinamento e notificação oficial da Vigilância Sanitária. Os resultados de qualidade da imagem foram comparados "antes" e "depois" da intervenção. RESULTADOS: O estudo demonstrou que dos 36 serviços que completaram a pesquisa, nenhum estava acima de 90 por cento de conformidade antes da intervenção. Após a intervenção, dez unidades atingiram mais de 90 por cento. As principais melhorias foram em relação aos chassis, compressão da mama e visualização de microcalcificações. CONCLUSÃO: Apesar de o Distrito Federal dispor de muitos serviços, na sua maioria não eram de qualidade. A intervenção foi eficaz para a melhoria da qualidade, porém, torna-se necessária uma ação continuada para resolver os problemas restantes e aumentar o impacto.


OBJECTIVE: The present study was aimed at evaluating the quality and the impact of an intervention involving inspection and education in mammography centers of the Federal District, Brazil. MATERIALS AND METHODS: Forty one mammography centers in the Federal District were studied in the period between 2000 and 2002. The intervention involved an initial inspection followed by a training activity and notification of mammography centers by the Federal District Sanitary Vigilance authority. The imaging quality was compared before and after the intervention. RESULTS: None of the 36 centers which completed the study reached more than 90 percent compliance with the standard imaging quality prior to the interventions, whereas ten were above 90 percent afterwards. Major improvements were observed in chassis maintenance, breast compression and visualization of microcalcifications. CONCLUSION: Despite the availability of a great number of mammography centers in the Federal District, most of them did not meet the required quality standards. The intervention has shown to be effective for improving the imaging quality, however a continued action is required to solve the remaining problems and increase the impact of the program.