Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
International Eye Science ; (12): 1849-1851, 2014.
Article in Chinese | WPRIM | ID: wpr-642032

ABSTRACT

AIM:To compare the refractive errors measured by the VISX WaveScan, OPD - Scan Ⅲ and the subjective refraction. METHODS: Seventy - six patients ( 152 eyes ) were recruited from January 2013 to December 2013. All patients were measured with subjective refraction by the phoropter (NIDEK, RT-5100), objective refraction by the WaveScan ( AMO Company, USA) , OPD-ScanⅢ ( Nidek Technologies, Japan). The sphere, cylinder, axis of the three methods were compared and analyzed. RESULTS: The sphere measured by WaveScan was lower than that by subjective refraction, the difference was 0. 13±0. 30D (t=3. 753, P0. 05). The value of the difference between WaveScan and subjective refraction was 5. 87o±6. 19o for the axis and the difference between OPD-Scan Ⅲ and subjective refraction was 3. 82o±3. 95o. There was statistic significance (t=2. 817, P=0. 006). CONCLUSION: For sphere and cylinder, WaveScan generated some deviation relative to subjective refraction. The Nidek OPD-ScanⅢ gives more accurate measures of objective refraction when compared with subjective refraction.