Rev. med. nucl. Alasbimn j ; 8(32)apr. 2006. ilus, tab
Article in English | LILACS | ID: lil-444091

ABSTRACT

Purpose: To compare with Picker’s thyroid phantom the contrast and the sensitivity obtained with ten cameras used for Tc-99m thyroid scintigraphy in a geographic area. Methods: The phantom image was acquired with the collimator and the acquisition parameters used in clinical practice. Contrast for the 12 mm hot (H12) and for the 9 and 12 mm cold (C9, C12) nodules was measured, as well as sensitivity (S in cp/MBq/s). Results: Measured minimum-maximum values were as follows. Pinhole cameras (6): S = 62-131; H12 = 0.66-0.77; C12 = 0.77-0.90; C9 = 0.59-0.76. Non-pinhole cameras (4): S = 55-85; H12 = 0.52-0.59; C12 = 0.62-0.75; C9 = 0.30-50. Conclusions: The pinhole collimators outclass the parallel hole ones in contrast. The sensitivity was the highest (S > 96) for five of the six pinhole collimators. As a direct consequence of this study, two centres decided to switch to pinhole collimator for thyroid scintigraphy.


Subject(s)
Humans , Radionuclide Imaging/instrumentation , Thyroid Gland , Radiometry/methods , Contrast Sensitivity , Gamma Cameras , Phantoms, Imaging , Sensitivity and Specificity , Technetium , Reference Values