1.
Journal of Biomedical Engineering ; (6): 1211-1213, 2009.
Article in Chinese | WPRIM | ID: wpr-244660

ABSTRACT

In this paper are analyzed the disadvantages of CGO-3000 digital subtraction angiography table Operating Box. The authors put forward a communication control scheme between single-chip microcomputer(SCM) and programmable logic controller(PLC). The details of hardware and software of communication are given.


Subject(s)
Humans , Angiography, Digital Subtraction , Equipment Design , Image Processing, Computer-Assisted , Microcomputers , Radiographic Image Enhancement , Software
2.
Chinese Medical Equipment Journal ; (6)1989.
Article in Chinese | WPRIM | ID: wpr-586283

ABSTRACT

Although the type of XHD150B-IO 800mA X-ray machine possesses a damage protection electric circuit for the main thyristors, with the first switch of photographic hand valve pressed, abnormal instant exposure still appears when double (in different phase) or more than three main thyristors are broken through, short-circuited or mis-energized, and consequently the high voltage element or X-ray machine will break down. Therefore the original electric circuit must be improved. Material: A few of leads. Method: We exchange the conjunction sequence of three NO (normal open) contacting points of radiograph high voltage preparing work relay--MAG SW- R with three reverse parallel connection of main thyristors. Conclusion: The normal procedure of radiographic exposure won't be disturbed after the improvement. When whichever one or more than two main thyristors is broken through, short-circuited or mis-energized, the spoilage can be detected instantly and the protection circuit works correspondingly, which guarantees the safety of X-ray machine.

SELECTION OF CITATIONS
SEARCH DETAIL