International Journal of Laboratory Medicine ; (12): 751-753, 2016.
Article in Chinese | WPRIM | ID: wpr-487156

ABSTRACT

Objective To implement the harmonization of thyroid-stimulating hormone(TSH) of Mindray assay system by par-ticipating in harmonization research program of International Federation of Clinical Chemistry (IFCC)-Standardization of Thyroid Function Tests(C-STFT ) .Methods A total of three combinations of TSH reagents and calibrators were used to measure 20 serum samples of healthy human .Within-run precision and batch-to-batch variation were assessed .The concept of harmonization was dem-onstrated by comparing our test results with All Procedure Trimmed Mean (APTM ) provided by IFCC and recalibrating with Mater Calibrators .Results The within-run precision was 1 .96% ,batch-to-batch variation was 0 .47% - 1 .15% ,depending on the level of TSH analyte .There existed a positive bias compared to APTM values .After recalibration with Mater Calibrators and Passing &Bablok regression ,the slope of method comparison was 1 .0 ,and correlation coefficient was more than 0 .975 .Conclusion By using a panel of real human specimen and recalibration based on APTM ,the test results of Mindray assay system could be harmonized with mainstream manufacturers globally .