1.
Journal of Biomedical Engineering ; (6): 886-893, 2023.
Article in Chinese | WPRIM | ID: wpr-1008913

ABSTRACT

In this paper, the differences between air probe and filled probe for measuring high-frequency dielectric properties of biological tissues are investigated based on the equivalent circuit model to provide a reference for the methodology of high-frequency measurement of biological tissue dielectric properties. Two types of probes were used to measure different concentrations of NaCl solution in the frequency band of 100 MHz-2 GHz. The results showed that the accuracy and reliability of the calculated results of the air probe were lower than that of the filled probe, especially the dielectric coefficient of the measured material, and the higher the concentration of NaCl solution, the higher the error. By laminating the probe terminal, liquid intrusion could be prevented, to a certain extent, to improve the accuracy of measurement. However, as the frequency decreased, the influence of the film on the measurement increased and the measurement accuracy decreased. The results of the study show that the air probe, despite its simple dimensional design and easy calibration, differs from the conventional equivalent circuit model in actual measurements, and the model needs to be re-corrected for actual use. The filled probe matches the equivalent circuit model better, and therefore has better measurement accuracy and reliability.


Subject(s)
Reproducibility of Results , Sodium Chloride , Calibration
SELECTION OF CITATIONS
SEARCH DETAIL