Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Acta Academiae Medicinae Sinicae ; (6): 17-20, 2017.
Article in English | WPRIM | ID: wpr-277905

ABSTRACT

Objective To evaluate the image quality and radiation dose of third-generation dual-source CT with tin filtration for spectral shaping and iterative reconstructions.Methods Thirty-five patients underwent low-dose CT (LDCT) for lung cancer screening on second-generation dual-source CT and follow-ups on third-generation dual-source CT. Image quality and radiation dose were compared between the two examinations.ResultsThe radiation dose of third-generation dual-source CT [dose-length product (DLP)(49.7±18.2)mGy·cm, effective dose (ED)(0.73±0.26)mSv] was lower than second-generation dual-source CT [DLP (86.37±13.44) mGy·cm, ED(1.20±0.42)mSv](t=6.01, P=0.000;t=6.57, P=0.000). The objective image noise of second-generation dual-source CT [(25.7±2.9)HU] was higher than that of third-generation dual-soure CT[(18.6±4.2)HU](t=5.24,P=0.000).The subjective image noise of second-generation dual-source CT [(4.60±0.49)scores] was significantly lower than that of third-generation dual-source CT [(4.80±0.40)scores] (t=4.15, P=0.000). Conclusion Chest CT for the detection of pulmonary nodules can be performed with third-generation dual-source CT that produces high image quality and low radiation dose when using a stellar infinity detector with spectral shaping.


Subject(s)
Humans , Early Detection of Cancer , Methods , Lung Neoplasms , Diagnostic Imaging , Radiation Dosage , Radiographic Image Interpretation, Computer-Assisted , Thorax , Diagnostic Imaging , Tomography, X-Ray Computed , Methods
SELECTION OF CITATIONS
SEARCH DETAIL