Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
International Journal of Laboratory Medicine ; (12): 2204-2205,2208, 2015.
Article in Chinese | WPRIM | ID: wpr-602323

ABSTRACT

Objective To evaluate analytical performance of homocysteine (Hcy) by circulating enzymatic method .Methods Re‐ferring to CLSI evaluation project and pertinent literature ,and by combining our actual works .we designed a verification procedure and experimental method .By using these above ,the precision ,accuracy ,analytical measurement range ,clinical reportable range of Hcy by circulating enzymatic method were evaluated .Results would be compared with the declaration of the manufacturer (NingBo Medical System Biotechnology Co .,Ltd) or desirable specifications derived from biologic variation .Results The results showed that the within‐run CV were 1 .26% and 0 .84% ,and the total CV were 1 .36% and 1 .32% ,less than 10% of the manufacturer′s statement .The relative bias between the results measured for calibrator at tow levels and target value was 3 .69% and 0 .69% ,less 10% .AMR was 3 .38-51 .81 μmol/L ,and the most suitable dilution rate was 1∶3 ,so the CRR was 3 .38-155 .43 μmol/L .Con‐clusion The analytical performance of Hcy analyzed by circulating enzymatic method is consistent with the standards which manu‐facturers has proclaimed ,so it is conform to the requirements of clinical .

SELECTION OF CITATIONS
SEARCH DETAIL