ABSTRACT

To evaluate flow-cytometry and Nageotte method for counting residual WBC in thrombocytaphoresis concentrates, their accuracies were determined by dilution studies separately; the repeatability was determined by measuring the interassay coefficient of variation for 14 replicates of a sample with known leukocyte concentration. 102 samples of leukocyte-depleted thrombocytaphoresis concentrates were detected by the above mentioned two methods, and the results were compared each other. The results showed that no difference was observed between two methods over a range of leukocyte concentrations from 0.8 to 10 WBC/microl (P > 0.05). In conclusion, flow-cytometry and Nageotte methods can be used for quality control of WBC-reduced blood components.