Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
1.
Korean Journal of Nuclear Medicine ; : 303-310, 2018.
Article in English | WPRIM | ID: wpr-786998

ABSTRACT

PURPOSE: This study aims to compare Monte Carlo-based codes' characteristics in the determination of the basic parameters of a high-resolution single photon emission computed tomography (HiReSPECT) scanner.METHODS: The geometry of this dual-head gamma camera equipped with a pixelated CsI(Na) scintillator and lead hexagonal hole collimator were accurately described in the GEANT4 Application for the Tomographic Emission (GATE), Monte Carlo N-particle extended (MCNP-X), and simulation of imaging nuclear detectors (SIMIND) codes.We implemented simulation procedures similar to the experimental test for calculation of the energy spectra, spatial resolution, and sensitivity of HiReSPECT by using 99mTc sources.RESULTS: The energy resolutions simulated by SIMIND, MCNP-X, and GATE were 17.53, 19.24, and 18.26%, respectively, while it was calculated at 19.15% in experimental test. The average spatial resolutions of the HiReSPECT camera at 2.5 cm from the collimator surface simulated by SIMIND, MCNP-X, and GATE were 3.18, 2.9, and 2.62 mm, respectively, while this parameter was reported at 2.82 mm in the experiment test. The sensitivities simulated by SIMIND, MCNP-X, and GATE were 1.44, 1.27, and 1.38 cps/µCi, respectively, on the collimator surface.CONCLUSIONS: Comparison between simulation and experimental results showed that among these MC codes, GATE enabled to accurately model realistic SPECT system and electromagnetic physical processes, but it required more time and hardware facilities to run simulations. SIMIND was the most flexible and user-friendly code to simulate a SPECT camera, but it had limitations in defining the non-conventional imaging device. The most important characteristics like time and speed of simulation, preciseness of results, and user-friendliness should be considered during simulations.


Subject(s)
Animals , Gamma Cameras , Magnets , Physical Phenomena , Tomography, Emission-Computed, Single-Photon
2.
Clinics ; 62(4): 377-384, 2007. ilus, tab
Article in English | LILACS | ID: lil-460018

ABSTRACT

PURPOSE: To evaluate the feasibility of using the Statistical Parametric Mapping (SPM) program for an automated, voxel-by-voxel assessment of regional cerebral blood flow (rCBF) deficits in Alzheimer's disease (AD) subjects relative to age-matched controls studied with a conventional, single-detector SPECT system. METHODS: We used a databank of 99mTc-HMPAO images of 19 patients with a diagnosis of probable AD and 15 elderly healthy volunteers; data were acquired using an Orbiter-Siemens single-detector SPECT system. Using SPM, images were transformed spatially, smoothed (12mm), and the data were compared on a voxel-by-voxel basis with t-tests. RESULTS: There were significant rCBF reductions in AD patients relative to controls involving regions predicted a priori to be affected in AD, namely the left temporal and parietal neocortices, and the right posterior cingulate gyrus (p<0.05, corrected for multiple comparisons). DISCUSSION: The location of rCBF reductions in AD subjects in our study is consistent with the deficits detected in previous functional imaging studies of AD using higher-resolution devices. This suggests the potential usefulness of using SPM for the analysis of data acquired with single-detector SPECT systems, despite the limited sensitivity and spatial resolution of such equipment.


OBJETIVO: Avaliar a viabilidade de emprego do programa Statistical Parametric Mapping (SPM) para investigar de forma automatizada, voxel-a-voxel, a presença de déficits de fluxo sanguíneo cerebral regional (FSCr) em pacientes com doença de Alzheimer (DA) comparados a sujeitos-controle pareados para idade, usando imagens de SPECT adquiridas com um equipamento convencional de detector único. MÉTODOS: Foi utilizado um banco de imagens adquiridas após injeção de 99mTc-HMPAO em 19 pacientes com diagnóstico provável de DA e 15 voluntários idosos saudáveis, usando um equipamento de SPECT Orbiter-Siemens de detector único. Empregando o programa SPM, as imagens foram transformadas espacialmente, suavizadas (12mm FWHM), e comparadas estatisticamente voxel-a-voxel entre os dois grupos, usando o teste de T. RESULTADOS: Foram identificadas reduções significativas de FSCr nos pacientes com DA comparados aos controles em regiões previstas a priori como afetadas por esta forma de demência, quais sejam os neocórtices temporal e parietal em hemisfério esquerdo e o cíngulo posterior direito (p<0,05, corrigido para comparações múltiplas). DISCUSSÃO: A localização dos focos de redução de FSCr em pacientes com DA no nosso estudo é, de forma geral, consistente com os achados de déficits cerebrais detectados em estudos anteriores de neuroimagem funcional na DA realizados com equipamentos de resolução espacial mais alta. Isto sugere o potencial de utilidade do programa SPM para a análise de dados de SPECT adquiridos com equipamentos de detector único, apesar da sensibilidade e resolução espacial limitadas de tais aparelhos.


Subject(s)
Aged , Female , Humans , Male , Alzheimer Disease/physiopathology , Brain Mapping/methods , Cerebrovascular Circulation/physiology , Tomography, Emission-Computed, Single-Photon/methods , Alzheimer Disease , Case-Control Studies , Data Interpretation, Statistical , Feasibility Studies , Radiopharmaceuticals
SELECTION OF CITATIONS
SEARCH DETAIL