Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Chinese Journal of Radiation Oncology ; (6): 165-168, 2014.
Article in Chinese | WPRIM | ID: wpr-443241

ABSTRACT

Objective To study the dosimetry parameters of 125I seed source (type Sinko BT-125-1) with thermoluminescent dosimeter (TLD) in the phantom.Methods The new type of phantom was modified to suit to measurement of a common type of 125I seed source.The AAPM TG43 protocol recommended measurements of dose-rate constant (Λ),radial dose function (gL (r)),and anisotropy function (F (r,θ)) have been performed in the phantom with TLD.Results The Λ was 0.928 cGyh-1 U-1.The gL(r) was determined at different radial distances r ranging from 1.0 to 10.0 cm with an interval of 1.0 cm ; and F (r,θ) at angles from 0° to 90° in 10° increments.The gL (r) of 125I seed source showed a difference of 9.6% at the most in comparison to the corresponding values of 125I seed source (type Amersham 6711).The difference in F(2 cm,θ) of 125I seed source and Amersham 6711 was up to 10.2% near the source end.With the phantom the combined standard uncertainty in the whole measurement was less than 6.0%.Conclusions The experimental results exhibit fairly small measurement uncertainties and good self-consistency.It's feasible to measure the dosimetry characters of permanent implant seeds in the modified phantom.

SELECTION OF CITATIONS
SEARCH DETAIL