1.
Indian J Ophthalmol ; 2010 Sept; 58(5): 440-442
Article in English | IMSEAR | ID: sea-136107

ABSTRACT

Glaucoma drainage devices are an option in refractory glaucomas for control of intraocular pressure (IOP). We evaluated the outcome of pars plana Ahmed glaucoma valve along with triamcinolone-assisted vitrectomy in 11 eyes with uncontrolled IOP on maximum tolerable antiglaucoma therapy. The mean preoperative IOP of 33.64 ± 5.99 (range 26 to 44 mmHg) decreased to 17.09 ± 2.26 (range 14 to 20 mmHg) and 17.45 ± 1.81mm of Hg (range 14 to 24 mmHg) at 6 and 12 months following surgery. The mean number of antiglaucoma medications decreased from 3.27 ± 0.05 to 0.64 ± 0.67 and 0.55 ± 0.6 at 6 and 12 months following surgery.


Subject(s)
Adolescent , Adult , Child , Glaucoma/physiopathology , Glaucoma/therapy , Glaucoma Drainage Implants , Glucocorticoids/therapeutic use , Humans , Intraocular Pressure , Male , Middle Aged , Triamcinolone/therapeutic use , Vitrectomy/methods , Young Adult