1.
Chinese Journal of Radiation Oncology ; (6): 142-145, 2009.
Article in Chinese | WPRIM | ID: wpr-396151

ABSTRACT

Objective To study the effect of CT scanning condition on the target position coordinate with WIMRT radiotherapy planning system using the" Z" form localizing frame. Methods The whole Head Phantom containing three metal targets was scanned under the following conditions: localizing frame transver-sally tilting, CT gantry tilting and localizing frame axially tilting. Then the three target coordinates were com-puted respectively and compared with those under the standard condition. Results The height and width of coordinate under the localizing frame transversally tilting condition deviated from those under the standard scanning condition,which was in an angle dependent manner. On the other hand ,a tendency was shown that the greater the absolute height value of the targets coordinate, the greater the width value deviation. And so was the relation between the absolute width value of the coordinate and height value deviation. If CT gantry angle was not zero degree or localizing frame axis is not in the same line with the CT bed moving direction, the calculated longititude value of the targets coordinate was mainly affected. Conclusions CT scanning must be performed strictly under the standard localizing scanning condition by a proficient physicist.

SELECTION OF CITATIONS
SEARCH DETAIL