ABSTRACT
A poly( GMA-EGDMA) coated SPME fiber was prepared using an in-situ polymerization by direct bonding to the surface of a polydopamine-modified stainless steel wire. Then the fiber was modified by sulfuric acid. A novel solid phase microextraction coating coupled to high performance liquid chromatography ( HPLC) method based on the as-prepared fiber was developed for the determination of four pharmaceuticals and personal care products ( PPCPs) in water samples. The influences of extraction parameters, including pH, extraction time, extraction temperature and salt addition were investigated. 3 mL water sample was extracted by the as-prepared fiber for 60 min at 30 ℃, and then desorbed with mobile phase for 30 min, respectively. Desorption solution was analyzed by HPLC-DAD ( diode array detection ) . The results indicated that the extraction yield of the fiber was good for four PPCPs. The linear correlation coefficients were>0. 997 with the linear range of 2-200 μg/L. The limits of detection (S/N=3) were 0. 5-5 μg/L with RSD (n=5) of 4. 1%-11. 9%. The recoveries of four PPCPs at spiked level of 20, 50, 100 μg/L were within the range of 70. 6%-105. 5%. The results showed that this method was easy, green, accurate and precise, and could be used to assay the four PPCPs in real water samples.