ABSTRACT
Objective:To analyze the dosimetric effects on off-center tumour treatment plan resulting from the MR-Linac-based isocenter position radiotherapy plan.Methods:The cases of 19 patients who were treated in Sun Yat-sen University Cancer Center in 2020 were collected in this study. Two different IMRT plans were designed for each patient with off-center tumor both for group A with planned isocenter position as IMRT and group B with planed target center position as geometric center. The conformity index and homogeneity index of target, the dose normal tissue and the number of MU were compared between two plans.Results:The two IMRT plans met clinical dosimetric requirements. No statistical differences were found both in homogeneity index and conformity index ( P>0.05). Also there was no differences found in doses to normal tissues. However, the MU number (1 149±903, t=2.804, P=0.012) in group A was higher than that in group B (970±652). Conclusions:It is feasible to perform MR-Linac-based off-center treatment plan.