1.
Korean Journal of Nuclear Medicine ; : 189-192, 1999.
Article in Korean | WPRIM | ID: wpr-186936

ABSTRACT

188Re (beta=22 MeV; gamma=155 keV; T1/2=16.9 hours) is an attractive therapeutic radioisotope which is produced from decay of reactor-produced tungsten-188 parent (T1/2=69 days). 188W has been produced from the double neutron capture reaction of 186W. 188Re can be easily obtained by elution of saline on alumina based 186W/188Re generator, which is commercially available. Complexes labelled with 188Re have been developed for the radiotherapy treatment of diseases because of the desirable nuclear properties of the radioisotope and it's chemical properties similar to those of technetium, a well established diagnostic agent.


Subject(s)
Humans , Aluminum Oxide , Neutrons , Parents , Radiotherapy , Technetium
2.
Korean Journal of Nuclear Medicine ; : 425-432, 1998.
Article in Korean | WPRIM | ID: wpr-130476

ABSTRACT

PURPOSE: For the purpose of using Re-188 as a therapeutic radionuclide, we performed the quality control of the W-188/Re-188 generation system. MATERIALS AND METHODS: Several quality control tests of the Re-188 eluate from generator were carried out of about 300 days. After elution of Re-188 with normal saline (20 ml), chromatogram and gamma-ray spectrum of Re-188 eluate were obtained. The presence of aluminum which was derived from the elumina bed of the generator was detected by using aluminum ion indicator kit. Re-188 eluate was allowed to decay for several days, and then W-188 breakthrough in the Re-188 eluate was measured by detecting gamma-ray at 227 keV and 290 keV. The pH and the pyrogenicity of the eluate were checked. The Re-188 eluate from the generator was 67.4+/-7.0% of W-188 during 270 days, and it was hightest at third day after previous elution. Radiochemical purity of Re-188 eluate obtained from chromatogram was higher than 99%. Gamma-ray spectrum of Re-188 eluate showed a peak at 155 keV. Aluminum ion and W-188 contamination were not detected. The pH of Re-188 eluate was 3 and the concentration yield was 85%. CONCLUSION: Our experiments and results on quality control tests of Re-188 eluate from W-188/Re-188 generator may be useful for setting W-188/Re-188 generator in hospitals.


Subject(s)
Aluminum , Hydrogen-Ion Concentration , Quality Control
3.
Korean Journal of Nuclear Medicine ; : 425-432, 1998.
Article in Korean | WPRIM | ID: wpr-130465

ABSTRACT

PURPOSE: For the purpose of using Re-188 as a therapeutic radionuclide, we performed the quality control of the W-188/Re-188 generation system. MATERIALS AND METHODS: Several quality control tests of the Re-188 eluate from generator were carried out of about 300 days. After elution of Re-188 with normal saline (20 ml), chromatogram and gamma-ray spectrum of Re-188 eluate were obtained. The presence of aluminum which was derived from the elumina bed of the generator was detected by using aluminum ion indicator kit. Re-188 eluate was allowed to decay for several days, and then W-188 breakthrough in the Re-188 eluate was measured by detecting gamma-ray at 227 keV and 290 keV. The pH and the pyrogenicity of the eluate were checked. The Re-188 eluate from the generator was 67.4+/-7.0% of W-188 during 270 days, and it was hightest at third day after previous elution. Radiochemical purity of Re-188 eluate obtained from chromatogram was higher than 99%. Gamma-ray spectrum of Re-188 eluate showed a peak at 155 keV. Aluminum ion and W-188 contamination were not detected. The pH of Re-188 eluate was 3 and the concentration yield was 85%. CONCLUSION: Our experiments and results on quality control tests of Re-188 eluate from W-188/Re-188 generator may be useful for setting W-188/Re-188 generator in hospitals.


Subject(s)
Aluminum , Hydrogen-Ion Concentration , Quality Control