Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add filters








Language
Year range
1.
Journal of the Korean Ophthalmological Society ; : 372-378, 2010.
Article in Korean | WPRIM | ID: wpr-155251

ABSTRACT

PURPOSE: To evaluate the repeatability of macular thickness measurements using time domain (TD) OCT and spectral domain (SD) OCT in diabetic macular edema. METHODS: In 42 eyes of 42 patients with diabetic macular edema, three consecutive macular measurements were performed with TD OCT and SD OCT, and measurements for macular thickness and total macular volume obtained by the two OCTs were compared. The within-subject standard deviation (Sw), coefficient of variation (CVw), and intraclass correlation coefficient (ICC) were calculated to assess repeatability, with agreement between measurements assessed with Bland Altman plots. The correlations were also evaluated via the Pearson's correlation coefficient. RESULTS: The Sw of TD OCT and SD OCT for foveal thickness, total macular volume were 29.67 micrometer/16.44 micrometer, 1.26 mm3/0.23 mm3, respectively, and were significantly lower in SD OCT. The ranges of the respective CVw and ICC values were 1.10-2.78%, 0.78~0.96 for TD OCT, and 0.29~0.94%, 0.92~0.99 for SD OCT. The SD OCT showed better repeatability for macular thickness measurements (all p< or =0.001). The 95% limits of agreement for foveal and total macular volume were 88.9 micrometer, 2.4 mm3, respectively. The Pearson's correlation coefficients of macular thickness and total macular volume between the two OCT methods were statistically significant (p=0.88-0.99). CONCLUSIONS: Although both OCTs proved reliable for macular thickness measurements in diabetic macular edema, SD OCT shows better repeatability than TD OCT. Although macular thickness measurements obtained from the two OCTs cannot be used interchangeably, there were statistically significant correlations between measurements obtained using the two OCTs.


Subject(s)
Humans , Eye , Macular Edema
2.
Journal of the Korean Ophthalmological Society ; : 710-716, 2009.
Article in Korean | WPRIM | ID: wpr-111142

ABSTRACT

PURPOSE:To compare the repeatability and agreement of macular thickness measurements using time domain (TD) optical coherence tomography (OCT) and spectral domain (SD) OCT in normal subjects. METHODS: Thirty-four normal subjects were included. Three consecutive macular measurements were taken with TD OCT and SD OCT. Total and regional macular thickness and total macular volume obtained by the two OCTs were compared. Within-subject standard deviation (Sw), coefficient of variation (CVw), and the intraclass correlation coefficient (ICC) were calculated to evaluate repeatability. The agreement was examined with Bland Altman plots. The correlation was also evaluated with Pearson correlation coefficients and multiple regression analysis. RESULTS: Sw for foveal thickness, total macular thickness, and total macular volume were 11.53 microm, 7.58 microm, and 0.48 mm3 for TD OCT and 6.69 microm, 2.55 microm, and 0.09 mm3 for SD OCT, respectively. The values for SD OCT were consistently lower, and this result was statistically significant. The range of the respective CVw and ICC values were 1.10~2.78% and 0.78~0.96% for TD OCT, and 0.29~0.94% and 0.92~0.99% for SD OCT, respectively. The SD OCT showed better repeatability for macular thickness measurements(all with p< or =0.001). The spans of 95% limits of agreement for foveal thickness, total macular thickness, and total macular volume were 67.94 microm, 29.01 microm, and 0.98 mm3, respectively. The Pearson correlation coefficient of foveal thickness, total macular thickness, and total macular volume between the two OCT's was statistically significant. CONCLUSIONS: Although both OCTs are reliable for macular thickness measurements, SD OCT shows better repeatability compared with TD OCT. Although macular thickness measurements obtained from the two OCTs can not be used interchangeably due to low agreement by different standards of measurement, there was a statistically significant correlation between the two OCT's.


Subject(s)
Tomography, Optical Coherence
SELECTION OF CITATIONS
SEARCH DETAIL