Chinese Journal of Radiological Medicine and Protection ; (12): 703-706, 2016.
Article in Chinese | WPRIM | ID: wpr-502306

ABSTRACT

Objective To establish the standard X/γ radiation fields with respect to the wrist dosimeter calibration curve as required by the international electrotechnial commission (IEC) technical specifications and the relevant national standard.Methods Air-kerma rate was determined in the X-ray beams,137Cs and 60Co radiation fields by a standard dosimeter.The wrist thermduminescent dosimetry (TLD) dosimeter was calibrated with personal dose equivalent values Hp (0.07) on the organization for standardization(ISO) wrist-phantom based on the radiation fields.Results The standard X/γ radiation field for the wrist dosimeter irradiation was established.Conclusions Established X/γ standard radiation field can be used for suc h technical services as wrist dosimeter calibration curve and energy response characteristics experiment.