Chinese Journal of Medical Instrumentation ; (6): 93-98, 2023.
Article in Chinese | WPRIM | ID: wpr-971311

ABSTRACT

Three different preclinical evaluation methods of MTF through-frequency response, MTF through-focus-response and expected visual acuity were used to compare and analyze the imaging differences of IOLs with four different optical designs. The research work could be used in the simultaneous vision IOLs in the optical design stage and verify the optical quality of the IOLs, the results can predict the visual representation of the patients better. The evaluation results can provide reference for IOL manufacturers and users in product design, development, validation and application selection.


Subject(s)
Humans , Prosthesis Design , Lenses, Intraocular , Vision, Ocular , Visual Acuity