Objective To explore the dilution regression method of coagulation detection(PT ,APTT) in fat blood samples . Methods We collected 40 normal blood coagulation specimens (no fat blood ,no jaundice ,no hemolysis) in Yan′an hospital of Kun-ming ,then we detected the PT and APTT of the original plasma and 3-fold diluted plasma and 5-fold diluted plasma ,the we used the data both of before dilution and diluted to do the linear regression analysis ,and finally we got the regression equations of each index .we also collected 33 fat blood samples in Yanan hospital of Kunming ,which be divide into three groups through the severity of triglycerides :mild fat blood group(1 .7 mmol/L≤TG0 .05) .Conclusion Dilution regression method can be used to detect the fat blood samples in the clinical coagulation detection .