Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Journal of Dental Rehabilitation and Applied Science ; : 269-277, 2017.
Article in English | WPRIM | ID: wpr-740462

ABSTRACT

PURPOSE: The aim of this study is to compare the performance of two piezoelectric engine systems by surveying satisfaction from dental clinicians. MATERIALS AND METHODS: Two piezoelectric systems were evaluated: TRAUS XUS10 (Saeshin), PIEZOSURGERY touch (Mectron). For this study, 20 dentists responded to the 11 questionnaires in which 5 point Likert-type scale was used. The two devices were operated for 10 seconds and measured 5 times to compare the maximum noise values. In heat emission test, the handpiece was operated for 3 minutes and heat was measured at three positions each. RESULTS: TRAUS XUS10 had higher satisfaction level on motor noise (P < 0.05). About function key and handpiece heat generation, PIEZOSURGERY touch showed higher satisfaction (P < 0.05) than TRAUS XUS10. The maximum noise level for each of the devices was confirmed to be 56.6 dB for the TRAUS XUS10 and 56.0 dB for PIEZOSURGERY touch. The two piezoelectric engines satisfied the safety standards with an operation temperature below 41℃ after having been operated for 3 minutes. CONCLUSION: Except for the function key and handpiece heat emission, TRAUS XUS10 has comparable performance with PIEZOSURGERY touch.


Subject(s)
Humans , Dentists , Hot Temperature , Noise , Piezosurgery
SELECTION OF CITATIONS
SEARCH DETAIL