Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
International Journal of Laboratory Medicine ; (12): 289-292, 2018.
Article in Chinese | WPRIM | ID: wpr-692662

ABSTRACT

Objective The ADVIA Centaur XP automatic chemiluminescence immunoassay system detec-tion performance of the thyroid hormones and antibodies for validation and evaluation.Methods With refer-ence to the American association of clinical laboratory standardization guide and other related documents,the ADVIA Centaur XP automatic chemiluminescence immunoassay system test 7 items thyroid hormones and an-tibodies(T3,T4,FT3,FT4,TSH,anti-TG,anti-TPO)of precision,accuracy,linear range and carry pollution rate for validation.Results Within the seven thyroid hormones and antibodies batch testing precision of CV and CV between batch precision between 1.51% -6.17% and 1.86% -6.17%.Is the average accuracy of bias of the largest testing project FT3(8.47%),but in the acceptable range,good correlation,correlation coefficient R of 0.994 or higher,Average bias <1/2 CLIA′88 TEa(12.5%).T3,T4,FT3,FT4,TSH,anti-TG and anti-T PO in 1.24-5.59 nmol/L respectively,and 60.07 -195.10 nmol/L,3.40 -22.87 pmol/L,14.59 -40.54 pmol/L,1.63-74.13 μIU/mL,60.10 -381.30 μIU/mL and 180.10 -531.50 μIU/mL range is linear,Carry pollution rate is between 0.04% -0.97%.Conclusion ADVIA Centaur XP automatic chemiluminescence im-munoassay system detecting thyroid hormones and antibodies results consistent with the data provided by the manufacturer,the test results are accurate and reliable,and can be used for the detection of clinical samples.

SELECTION OF CITATIONS
SEARCH DETAIL