Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add filters








Year range
1.
J. health med. sci. (Print) ; 8(1): 45-50, ene.-mar. 2022. ilus, graf, tab
Article in English | LILACS | ID: biblio-1395758

ABSTRACT

Radiation absorbed doses to organs outside the radiation therapy treatment beam can be significant and therefore of clinical interest. Two sets of out-of-beam measurements were performed measuring the leak dose and the scattered dose, at 5 points within the accelerator components (accelerator tube and collimator) and at 21 points on the equipment and surroundings based on a positioning scheme. For this purpose, 52 Optically Stimulated Luminescence (OSL) dosimeters were used in a latest generation helical linear accelerator. Of the 200 cGy fired at a cheese-like phantom, 0.332% of the out-of-beam dose contribution was found to come from the leak and 0.784% was transformed into scattering. For these dose values, estimates of the risk of second tumors in long-term survivors indicate a reduced probability of acquiring a second secondary radiation malignancy, based on information from the 1990 BEIR Committee report.


La dosis absorbida de radiación a órganos fuera del haz de tratamiento de radioterapia puede ser significativa y, por lo tanto, de interés clínico. Se realizaron dos sets de mediciones fuera del haz para determinar la dosis de fuga y la dosis dispersa, en 5 puntos dentro de los componentes del acelerador (tubo de aceleración y colimador) y 21 puntos en el equipo y alrededores basado en un esquema de posicionamiento. Para este fin se utilizaron 52 dosímetros de luminiscencia estimulada ópticamente (OSL, Optically Stimulated Luminescence), en un acelerador lineal helicoidal de última generación. De los 200 cGy disparados a un maniquí tipo queso, se encontró que el 0.332% de la contribución de dosis fuera del haz provenía de la fuga y 0.784% se transforma en dispersión. Para estos valores de dosis, las estimaciones del riesgo de segundos tumores en los supervivientes a largo plazo indican una reducida probabilidad de contraer una segunda malignidad por radiación secundaria, según la información del informe del Comité BEIR de 1990.


Subject(s)
Humans , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Intensity-Modulated/methods , Optically Stimulated Luminescence Dosimetry , Radiometry/instrumentation , Thermoluminescent Dosimetry , Calibration , Luminescence , Luminescent Measurements
2.
Rev. chil. radiol ; 22(2): 70-75, jun. 2016. ilus, tab
Article in Spanish | LILACS | ID: lil-796827

ABSTRACT

Abstract. There are several methods for measuring the absorbed radiation dose of occupationally exposed professionals, but it is unknown if these are reproducible at low doses. The aim of this study is to determine the level of reproducibility of thermoluminescent dosimeters (TLD) and optically stimulated luminescence dosimeters (OSL) in simulated conditions. A reproducibility study was performed using two types of dosimeters, which were placed in the location of medical interventionist work area inside the catheter laboratory. An anthropomorphic phantom was used as a patient and the same projections, collimation and radiographic techniques that were employed in the past 30 angiographies, an examination considered as low exposure in the context of interventional cardiology. Lin's concordance coefficient correlation was calculated, obtaining a value of -0.006 with a 95% CI of -0.069 to 0.056 for TLD dosimeters with OSL. The data obtained showed poor reproducibility at low doses of radiation.


Resumen. Existen varios métodos para medir la dosis de radiación absorbida por los profesionales ocupacionalmente expuestos, pero se desconoce si estos son reproducibles entre sí a bajas dosis. El objetivo del estudio fue determinar el grado de reproducibilidad entre los dosímetros termoluminiscentes (TLD) y los dosímetros luminiscentes ópticamente estimulados (OSL), en condiciones simuladas. Se realizó un estudio de reproducibilidad, empleando 2 tipos de dosímetros, los cuales se dispusieron en la ubicación de trabajo del médico intervencionista dentro del pabellón. Se empleó un fantoma antropomórfico como paciente y se replicaron las mismas proyecciones, colimación y técnicas radiológicas empleadas en las últimas 30 coronariografías, examen considerado de baja exposición en el contexto de cardiología intervencionista. Se aplicó el coeficiente de correlación de concordancia de Lin, obteniéndose un valor de -0,006 con un IC al 95% de -0,069 a 0,056 para los dosímetros TLD con OSL. Los datos obtenidos mostraron una escasa reproducibilidad a bajas dosis de radiación.


Subject(s)
Humans , Radiation Dosage , Cardiology , Radiation Monitoring/methods , Radiation Monitoring/instrumentation , Reproducibility of Results , Radiation Dosimeters , Optically Stimulated Luminescence Dosimetry
SELECTION OF CITATIONS
SEARCH DETAIL