Chinese Journal of Medical Instrumentation ; (6): 5-6, 2005.
Artículo en Chino | WPRIM | ID: wpr-241116

RESUMEN

Some research on intelligent non-invasive temporary pacemakers is introduced in this paper. An industrial computer, some IC chips and other elements are used to construct its hardware, and its software is in C++ language. The experimental device has some intelligent functions of recognizing some arrhythmia. The system has a pacemaker module and an ECG monitor module. Its software includes a main program, a RS-232C communication program, a printer VxD, a pacing control VxD and ECG signal pretreatment and recognizing program and so on. The pacing-generating circuit is employed to make the precision control of pacing current. The communication between industrial-computer system and ECG module is completed through the DLL. The real time processing of ECG signals is based on filter method for a higher recognizing ratio. The system calculates several parameters to recognize certain arrhythmia and uses MIT/BIH database to validate the reliability of ECG recognition.


Asunto(s)
Humanos , Algoritmos , Arritmias Cardíacas , Diagnóstico , Terapéutica , Inteligencia Artificial , Ecocardiografía , Diseño de Equipo , Marcapaso Artificial , Clasificación , Procesamiento de Señales Asistido por Computador , Diseño de Software
2.
Chinese Medical Equipment Journal ; (6)2003.
Artículo en Chino | WPRIM | ID: wpr-684378

RESUMEN

This paper presents a high voltage generator for the defibrillator.Its principle and design method are introduced.The high voltage generator is the core part of the defibrillator which is necessary toany hospital.There is nodomestic-patented defibrillator by now and published paper specialized in the principle and design method of the high voltage generator for the defibrillator.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA