1.
Rev. dent. press ortodon. ortopedi. facial ; 13(4): 60-66, jul.-ago. 2008. ilus
Artículo en Portugués | LILACS | ID: lil-487765

RESUMEN

OBJETIVO: avaliar a efetividade do esterilizador com esferas de vidro Steri® 350, quanto ao controle de infecção das partes ativas dos alicates ortodônticos. METODOLOGIA: foram utilizados nove alicates ortodônticos, previamente esterilizados em autoclave à temperatura de 121ºC, durante 20 minutos. Posteriormente, as partes ativas dos alicates foram contaminadas com Bacillus stearothermophilus. Logo após, as pontas dos alicates foram colocadas no Steri® 350, durante os períodos de 3, 5, 10, 15, 20, 30 e 40 segundos, em temperatura de 255ºC, para avaliar a eficácia da esterilização. RESULTADOS E CONCLUSÕES: o esterilizador com esferas de vidro mostrou-se eficaz no controle do crescimento de Bacillus stearothermophilus nas partes ativas dos alicates ortodônticos, a partir de 10 segundos de exposição à temperatura de 255ºC.


AIM: The aim of this research was to evaluate the efficiency of Steri® 350glass bead sterilizer, on the infection control of the active part of the orthodontic pliers. METHODS: Nine orthodontic pliers were sterilized in the steam autoclave at 121ºC in 20 minutes and subsequently the active part of each plier was dipped in a culture for Bacillus stearothermophilus. Later, the active part of each plier was putted into the Steri® 350 in the following periods: 3, 5, 10, 15, 20, 30, and 40 seconds at 255ºC to evaluate the efficiency of this method in the infection control. RESULTS AND CONCLUSIONS: This sterilization method has been effective to control the presence of Bacillus stearothermophilus on the active part of orthodontic pliers at 255ºC in 10 seconds.


Asunto(s)
Esterilización/métodos , Ortodoncia/instrumentación , Geobacillus stearothermophilus , Infecciones
2.
Acta odontol. venez ; 45(1): 29-32, 2007. tab
Artículo en Español | LILACS | ID: lil-483978

RESUMEN

El objetivo de este estudio fue evaluar la eficacia de agentes de desinfección indicados para polisulfuros (mercaptanos), poliéteres y siliconas por condensación y por adición. Fueron confeccionadas noventa muestras de cada material, siendo treinta de ellas contaminadas con Streptococcus mutans o Staphylococcus aureus o Candida albicans. De cada solución microbiana fueron retiradas diez muestras del polisulfeto y de ambas las siliconas, que fueron inmersas por diez minutos en glutaraldehído al 2 por ciento y otras diez inmersas en agua destilada estéril (control negativo). Después de un nuevo lavado en agua destilada, las muestras fueron transferidas a medios de cultivo estériles. Las diez muestras restantes no fueron sometidas al agente de desinfección y fueron transferidas a medios asociados a agentes antimicrobianos específicos (control positivo). La turbidez de los medios de cultivo fue evaluada como indicativo del crecimiento microbiano siguiéndose a la incubación por 24 h a 37O C y se realizó la dilución y sembrado en placas de Petri para contar las colonias. Para el poliéter fue ejecutado el mismo procedimiento, pero el agente de desinfección usado fue el hipoclorito de sodio a 1. No hubo turbidez comprobatoria del crecimiento microbiano en ninguno de los medios de cultivo que contenían los especimenes sometidos a los agentes de desinfección. Se concluyó que el glutaraldehído al 2 por ciento es un agente de desinfección eficaz para el polisulfuro y para las siliconas por adición y por condensación, así como el hipoclorito a 1 por ciento es eficaz para el poliéter, para los microorganismos evaluados


The aim of this in vitro study was to evaluate the disinfection efficacy of elastomeric impression materials such as polysulfides, polyethers, condensation and addition silicones. Ninety samples of each material were made and every thirty samples contaminated with Streptococcus mutans or Staphylococcus aureus or Candida albicans. From each microbial solution there were taken ten samples of polysulfide and of both silicones which were immersed for ten minutes in a 2% glutaraldehyde solution and other ten immersed in sterile distilled water (negative control). After being washed again in distilled water, the samples were transferred to sterile culture medium. The last ten samples were not submitted to any disinfection agent and were transferred to mediums associated with specific anti-microbial agents (positive control). The turbidity of the culture mediums was evaluated as an indication of the microbial growth after a period of 24-hour incubation at 37oC and the dilution and seeding in Petri dishes were done to count the colonies. The same procedure was done for the polyether, but the disinfection agent used was 1% sodium hypoclorite. There was no proved turbidity of the microbial growth in any of the culture mediums which had samples submitted to the disinfection agents tested. It was concluded that the 2% glutaraldehyde is an effective disinfection agent to polysulfides and to both condensation and addition silicones, as well as 1% sodium hypoclorite is effective to polyether, for the tested microorganisms.


O objetivo deste estudo in vitro foi avaliar a eficácia de agentes de desinfecção indicados para polissulfetos, poliéteres e siliconas de condensação e adição. Foram confeccionadas noventa amostras de cada material, sendo cada trinta contaminadas com Streptococcus mutans, Staphylococcus aureus ou Candida albicans. De cada solução microbiana foram retiradas dez amostras do polissulfeto e de ambas as siliconas, que foram imersas por dez minutos no glutaraldeído a 2% e outras dez imersas em água destilada estéril (controle negativo). Após nova lavagem em água destilada, as amostras foram transferidas para meios de cultura estéreis. As dez amostras restantes não foram submetidas ao agente de desinfecção e foram transferidas para meios associados a agentes antimicrobianos específicos (controle positivo). A turbidez dos meios de cultura foi avaliada como indicativo de crescimento microbiano após incubação por 24 h a 37O C e realizou-se a diluição e semeadura nas placas de Petri para contagem de colônias. Para o poliéter foi executado o mesmo procedimento, porém o agente de desinfecção empregado foi o hipoclorito de sódio a 1%. Não houve turvação comprobatória de crescimento bacteriano em nenhum dos meios da cultura que continham corpos de prova submetidos aos agentes de desinfecção. Concluiu-se que o glutaraldeído a 2% é um agente de desinfecção eficaz para o polissulfeto e para as siliconas por adição e por condensação, assim como o hipoclorito a 1% é eficaz para o poliéter, para os microorganismos testados.


Asunto(s)
Control de Infección Dental/instrumentación , Desinfectantes/clasificación , Glutaral/química , Hipoclorito de Sodio/química , Materiales de Impresión Dental/normas , Elastómeros de Silicona , American Dental Association , Protocolos Clínicos , Recuento de Colonia Microbiana , Medios de Cultivo , Candida albicans/aislamiento & purificación , Poliésteres , Sulfuros , Staphylococcus aureus/aislamiento & purificación , Streptococcus mutans/aislamiento & purificación