Chinese Journal of Medical Instrumentation ; (6): 283-286, 2022.
Artículo en Chino | WPRIM | ID: wpr-928905

RESUMEN

An integrated digital magnetic resonance imaging spectrometer was proposed in this study. By using the FPGA chip Artix7 200T, timing control, data processing, digital frequency conversion and phase control were implemented into a single-chip, thus effectively improved timing accuracy and phase accuracy, while avoided the structural design complexity caused by multi-board connection and improved system integration and imaging quality.


Asunto(s)
Diseño de Equipo , Imagen por Resonancia Magnética