1.
Chinese Journal of Analytical Chemistry ; (12): 189-191, 2001.
Artículo en Chino | WPRIM | ID: wpr-411178

RESUMEN

A new flow-injection chemiluminescence method for the determinstion of trace amounts of sulfide was developed. It is based on the catalysis of sulfide to chemilumincence reacion between luminol and hydrogen peroxid. The linear range extends from 2.0 × 10-9 to 8.0 × 10-7 g/mL and the detection limit is 5.6 × 10-10 g/mL.The relative standard deviation (RSD) is less than 5% for 1.0 × 10-7 g/mL sulfide solution in eleven repeated measurements. The method has the advantages of simplicity, rapidity, sensitivity and has also been applied to the determination of trace amounts of sulfide in water samples.

2.
Chinese Journal of Analytical Chemistry ; (12): 296-298, 2001.
Artículo en Chino | WPRIM | ID: wpr-410765

RESUMEN

The chemiluminescence reaction of amidopyrine-potassium permanganate with formaldehyde as an enhancer was investigated by flow injection system. A method for the determination of amidopyrine on the basis of this technique was proposed. The detection limit is 3.0×10-8 g/mL, the relative standard deviation is 1.3% (4.0×10-6 g/mL amidopyrine,n=11).The linear range is 1.0×10-7~8.0×10-5g/mL amidopyrine. The method has been applied to the determination of amidopyrine in the antondin injection.

SELECCIÓN DE REFERENCIAS