Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Añadir filtros








Intervalo de año
1.
Journal of Interventional Radiology ; (12): 857-860, 2015.
Artículo en Chino | WPRIM | ID: wpr-481181

RESUMEN

Objective To investigate the feasibility of centerline measurement method in estimating aortic diameter at the proximal landing zone in Stanford B type aortic dissection. Methods CT angiography materials of 30 patients with type B aortic dissection were randomly selected from the hospital database (24 males with a median age of 49.5 years), which were retrospectively analyzed with multiplanar reformation (MPR) and centerline technique by two experts in vascular radiology. Difference between two measurement techniques was analyzed by using mixed linear model, and the agreement of measurements between two readers as well as between two techniques were evaluated by Bland-Altman plots. Results The diameters measured with MPR method by two experts were (29.73±2.99) mm and (29.86±2.95) mm respectively, while the diameters measured with centerline measurement method by two experts were (29.66 ±2.81) mm and (29.71 ±2.91) mm respectively. No statistically significant differences in the diameter value existed between the two measurement methods, although the results determined by centerline measurement method were more stable. Conclusion In determining aortic diameter at the proximal landing zone in Stanford B type aortic dissection, the centerline analysis provides a checking method for MPR measurement.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA