1.
Indian J Med Sci ; 2007 Feb; 61(2): 83-90
Artículo en Inglés | IMSEAR | ID: sea-68375

RESUMEN

BACKGROUND: There is widespread concern about radiation doses imparted to patients during cardiology procedures in the medical community. The current study intends to audit and optimize radiation dose to patients undergoing coronary angiography (CA) performed using two dedicated cardiovascular machines. MATERIALS AND METHODS: One hundred and forty nine patients who underwent CA are reported in this study. Dose auditing was done by implementing dose reduction strategies using spectral filters and by evaluating work practices of operators involved in performing CA. STATISTICAL ANALYSIS: A Student's 't' test was used to analyze the statistical significance. RESULTS AND CONCLUSION: The radiation dose imparted to patients was measured using dose area product (DAP) meter. The mean DAP values during CA before optimization was 55.86 Gy cm2 and after optimization was 27.71 Gy cm2. No ill-effects of radiation were reported for patients who underwent CA. Use of copper filtration may be recommended for procedures performed using cardiovascular machines.


Asunto(s)
Adulto , Anciano , Protocolos Clínicos , Angiografía Coronaria/métodos , Femenino , Fluoroscopía/métodos , Humanos , Masculino , Persona de Mediana Edad , Dosis de Radiación , Monitoreo de Radiación , Protección Radiológica/métodos , Efectividad Biológica Relativa
2.
Indian J Med Sci ; 2005 Dec; 59(12): 527-33
Artículo en Inglés | IMSEAR | ID: sea-66636

RESUMEN

BACKGROUND: Abdominal embolization procedures performed using digital subtraction angiography (DSA) is on the increase in the present-day scenario owing to their diagnostic and therapeutic values. These procedures involve prolonged fluoroscopy times and may tend to impart high radiation dose to patients if adequate radiation safety measures are not taken. AIM: To evaluate radiation dose imparted to patients and the work practices involved therein during abdominal embolization procedures. MATERIALS AND METHODS: Forty-two patients who underwent abdominal embolizations performed using DSA equipment were included in the study. Dose area product (DAP) was measured using DAP meter and values obtained were used for calculating entrance surface dose (ESD). Work practices of personnel involved in conducting the procedure were evaluated based on the choice of field sizes, selection of appropriate fluoro-modes, and optimization techniques. RESULTS AND CONCLUSIONS: The mean ESD values during hepatic embolization, renal embolization, splenic artery embolization and transarterial chemoembolization (TACE) were 1.2, 1.01, 1.19, and 1.03, respectively. No deterministic effects of radiation, such as transient or main erythema, were noticed for a few patients whose doses exceeded the threshold doses.


Asunto(s)
Abdomen , Angiografía de Substracción Digital , Quimioembolización Terapéutica , Embolización Terapéutica , Femenino , Fluoroscopía , Humanos , Masculino , Dosis de Radiación , Monitoreo de Radiación/métodos , Radiografía Intervencional , Piel/efectos de la radiación