J.health med.sci. ; 9(3): 3-9, jul.2023. ilus, tab
Artículo en Inglés | LILACS | ID: biblio-1519661

RESUMEN

ABSTRACT: The aim of this work is to provide a methodology for evaluating the committed effective dose E(50) due to the incorporation of [18F] FDG in the occupationally exposed worker (OEW) of the Cyclotron-PET/CT Laboratory of the Centro de Investigación en Ciencias Atómicas, Nucleares y Moleculares (CICANUM) at Universidad de Costa Rica using in vivo measurements. The measurement system was calibrated to perform in vivo measurements and defined as the corresponding bioassay function for the radiopharmaceutical used. The conversion factor was assessed with a known activity of 18F in the geometry and measurement time established. Among the most relevant results, the measurement parameters and the calibration procedure were defined. A value of 1.73 x 103 Bq/cps for in vivo brain measurements was obtained as a conversion factor. This study provides a methodology, to evaluate the committed effective dose due to the incorporation of 18F-FDG in a radionuclide production and diagnostic center


Asunto(s)
Protección Radiológica , Exposición Profesional/efectos adversos , Ciclotrones/instrumentación , Dosis de Radiación