J Cancer Res Ther ; 2008 Oct-Dec; 4(4): 173-7
Artículo en Inglés | IMSEAR | ID: sea-111400

RESUMEN

AIM: To evolve a fast dose verification method for high-dose-rate (HDR) brachytherapy treatment plans and to demonstrate its applicability in different clinical cases. MATERIALS AND METHODS: We developed a software tool in VC++ for the Varisource HDR unit for HDR dosimetry plan verification using TG-43 parameters. HDR treatment dosimetry of a number clinical cases using Varisource was verified by comparison with the treatment planning system (TPS). RESULTS: A number of different types of clinical cases treated by Varisource were evaluated. TPS calculated dose values and verification code calculated dose values were found to agree to within 3% for most of the dose calculation points. CONCLUSIONS: We have validated with clinical cases a fast and independent dose verification method of the dosimetry at selected points for HDR brachytherapy treatments plan using TG-43 parameters. This can be used for the verification of the TPS calculated dose at various points. The code is written to work with Varisource, but it can conceivably be modified for other sources also by using the fitted constant of the respective source.


Asunto(s)
Anisotropía , Braquiterapia/métodos , Humanos , Modelos Estadísticos , Método de Montecarlo , Probabilidad , Lenguajes de Programación , Radiometría/métodos , Dosificación Radioterapéutica , Planificación de la Radioterapia Asistida por Computador/métodos , Radioterapia de Alta Energía/métodos , Programas Informáticos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA