Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Añadir filtros








Intervalo de año
1.
International Journal of Laboratory Medicine ; (12): 613-614, 2015.
Artículo en Chino | WPRIM | ID: wpr-461470

RESUMEN

Objective To use the AU2700 biochemical analyzer to evaluate the performance of the reagent kit for determining se-rum glycated albumin(GA)by the enzymatic method.Methods On the basis of the relevant documents of CLSI-EP,the perform-ance evaluation of the Lucica GA-L reagent kit for detecting GA was performed on the precision,accuracy,linearity range and refer-ence range;the correlation between the results detected by the two kinds of reagent kit (enzymatic method)was analyzed.Results The reagent had high precision(CV<3%);the dilution had good linearity;the average recovery rate of GA was 97.7%,the average recovery rate of albumin(ALB)was 96.43%;the detection results by the two kinds of reagent kit had good correlation (r2 =0.976, P <0.01).Conclusion The Lucica GA-L reagent kit shows the good analytical performance on the AU2700 biochemical analyzer, which is consistent to the manufacturers statement and can be better used in clinic.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA