Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Añadir filtros








Intervalo de año
1.
Indian Pediatr ; 2022 Aug; 59(8): 620-622
Artículo | IMSEAR | ID: sea-225360

RESUMEN

Objective: To compare the continuous real time blood glucose (CG) measurement with venous laboratory blood glucose (LG) level in neonates during perioperative period. Methods: Glucose levels were measured simultaneously by CG, glucometer glucose (GG) and LG at 40 time points in ten neonates during perioperative period. Intraclass correlation coefficient (ICC) and Bland Altman analysis were used for comparison. Result: Correlation between CG and LG was excellent (ICC= 0.953; P<0.001), and average difference was 23.8 (95%CI 52.9 to -5.3) mg/dL, showed better reliability than at hyperglycemic state (ICC=0.653; P=0.006). The GG-LG showed excellent reliability with ICC = 0.985; P<0.001 and average difference of 15.4 (95% CI 30.7 to 0.1) mg/dL. CG at euglycemic state (ICC= 0.880; P<0.001). Conclusion: CG measurement is reliable for blood sugar estimation in neonates; but has lower reliability for hyperglycemia. The continuous trend of glucose measurement by CG is helpful for timely diagnosis of hyperglycemia during perioperative period in neonates.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA