RESUMEN
PURPOSE: This treatment planning study was undertaken to evaluate the impact of beam angle configuration of intensity-modulated radiotherapy (IMRT) on the dose of the normal liver in hepatocellular carcinoma (HCC). MATERIALS AND METHODS: The computed tomography datasets of 25 patients treated with IMRT for HCC were selected. Two IMRT plans using five beams were made in each patient; beams with equidistance of 72degrees (Plan I), and beams with a 30degrees angle of separation entering the body near the tumor (Plan II). Both plans were generated using the same constraints in each patient. Conformity index (CI), homogeneity index (HI), gamma index, mean dose of the normal liver (Dmean_NL), Dmean_NL difference between the two plans, and percentage normal liver volumes receiving at least 10, 20, and 30 Gy (V10, V20, and V30) were evaluated and compared. RESULTS: Dmean_NL, V10, and V20 were significantly better for Plan II. The Dmean_NL was significantly lower for peripheral (p = 0.001) and central tumors (p = 0.034). Dmean_NL differences between the two plans increased in proportion to gross tumor volume to normal liver volume ratios (p = 0.002). CI, HI, and gamma indices were not significantly different for the two plans. CONCLUSION: The IMRT plan based on beams with narrow separations reduced the irradiated dose of the normal liver, which would allow radiation dose escalation for HCC.
Asunto(s)
Humanos , Carcinoma Hepatocelular , Hígado , Radioterapia de Intensidad Modulada , Carga TumoralRESUMEN
The pencil beam convolution (PBC) algorithms in radiation treatment planning system have been widely used to calculate the radiation dose. A new photon dose calculation algorithm, referred to as the anisotropic analytical algorithm (AAA), was released for use by the Varian medical system. The aim of this paper was to investigate the difference in dose calculation between the AAA and PBC algorithm using the intensity modulated radiation therapy (IMRT) plan for lung cancer cases that were inhomogeneous in the low density. We quantitatively analyzed the differences in dose using the eclipse planning system (Varian Medical System, Palo Alto, CA) and I'mRT matirxx (IBA, Schwarzenbruck, Germany) equipment to compare the gamma evaluation. 11 patients with lung cancer at various sites were used in this study. We also used the TLD-100 (LiF) to measure the differences in dose between the calculated dose and measured dose in the Alderson Rando phantom. The maximum, mean, minimum dose for the normal tissue did not change significantly. But the volume of the PTV covered by the 95% isodose curve was decreased by 6% in the lung due to the difference in the algorithms. The difference dose between the calculated dose by the PBC algorithms and AAA algorithms and the measured dose with TLD-100 (LiF) in the Alderson Rando phantom was -4.6% and -2.7% respectively. Based on the results of this study, the treatment plan calculated using the AAA algorithms is more accurate in lung sites with a low density when compared to the treatment plan calculated using the PBC algorithms.
Asunto(s)
Humanos , Pulmón , Neoplasias PulmonaresRESUMEN
The energy spectra for electron beam of medical linear accelerator were calculated using a GEANT4 Medical Linac 2 example code. The incident electron mean energy were 6, 9, 12, 16, 20 MeV. This code was designed to calculate electron beam energy spectra according to material, thickness and location of electron scattering foil affecting electron beam characteristic. Lead, Copper, Aluminum and Gold were used for scattering foil. The energy distribution for electron and photon were analyzed by changing position of scattering foil in the head of linear accelerator. The effect of electron scattering foil on energy spectra which is basic data of simulation for medical linear accelerator were presented. The calculated results would be used in design of medical accelerator head.
Asunto(s)
Aluminio , Cobre , Electrones , Cabeza , Aceleradores de PartículasRESUMEN
In this study, Geant4 based Monte Carlo simulations were carried out for medical linear accelerator. Modified Medical Linac2 toolkit was used for calculation. The energy spectrum, most probable energy and the photon mean energy compared with the published results using the EGS4 code. The results well agreed with published results. The calculated results of photon fluence, energy fluence and mean energy according to the radius from the centre of the beam were analyzed. Monte Carlo simulation using Medical Linac2 code is considered to be useful for analysis of medical linear accelerator. Because the calculated results varies depending on Physics List model for same head structure. It it important to choose the right model for research purpose. Monte Carlo simulation using GEANT4 Medical Linac2 is a valuable for any novice to adopt this code to the study related to 6 MV photon fluence from medical linear accelerator.