Laboratory Medicine Online ; : 207-213, 2020.
Artículo | WPRIM | ID: wpr-836923

RESUMEN

Background@#This study was conducted to evaluate the analytical performance of the SelexOnTM B-type natriuretic peptide (BNP) assay (Osang Healthcare Inc., Korea), a new rapid lateral flow immunoassay for point of care (POC) testing using whole blood. @*Methods@#The imprecision, linearity, and method comparison of SelexOnTM BNP assay were evaluated. Two commercial BNP assays, the ADVIA Centaur® BNP (Siemens Health Care diagnostics Inc., USA) and the Triage® BNP assays (Alere, USA), were included for method comparison using 100 whole blood samples from patients. The reference interval was verified using 120 residual samples from health examination participants. @*Results@#The SelexOn BNP had total CVs of 20.3%, 13.3%, and 10.3% in BNP concentrations of 89.44 pg/mL, 480.71 pg/mL, and 1,201.84 pg/mL of control materials, respectively. Linearity was observed from 56 pg/mL to 1544 pg/mL. The SelexOn BNP (y) regression equation was y=0.9706x-21.68 with Centaur BNP (x) (r=0.930) and y=0.7600x+0.0506 with Triage BNP (x) (r=0.845), respectively. The predicted mean difference (%) of the SelexOn BNP at the clinical decision levels (100 pg/mL) was up to 25% lower than the two comparative methods. The SelexOn BNP levels were below 50 pg/mL in 114 (95%) of the 120 samples. @*Conclusions@#The SelexOn BNP using EDTA was developed as a POC test for differential diagnosis or treatment monitoring for acute heart failure. However, clinical decision values must be improved to be compatible with other BNP methods.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA