Chinese Journal of Analytical Chemistry ; (12): 779-780, 2014.
Artículo en Chino | WPRIM | ID: wpr-671750

RESUMEN

A method for the simultaneous determination of oxygen and nitrogen in synthetic diamonds by inert gas high temperature extraction-impulse heating method was proposed. The sample weight, the selection of analysis power and the calibration curves of oxygen and nitrogen were discussed. Oxygen and nitrogen in analytical samples are determined. Values of RSDs (n=7) for oxygen and nitrogen were less than 4. 5% and 4. 0% respectively. The analytical results of oxygen and nitrogen obtained by the proposed method were in good agreement with those by inert gas fusion-impulse heating method.

2.
Chinese Journal of Analytical Chemistry ; (12): 457-458, 2014.
Artículo en Chino | WPRIM | ID: wpr-671730

RESUMEN

A method for the determination of Fe, Co, Mn and Ni in synthetic diamonds by inductively coupled plasma atomic emission spectrometry ( ICP-AES) was proposed. The synthetic diamond sample was decomposed completely, while the sample was burned in air at 1000 ℃ for 10 h, and then a mixed acid of H2 SO4 , aqua regia and HClO4 was used for the dissolving the residue of the sample. In this method, the limits of detection of Fe, Co, Mn and Ni were 0. 0147, 0. 0018, 0. 0006 and 0. 0027 mg/L, respectively. Under the optimum condition, Fe, Co, Mn and Ni in synthetic diamond sample were determined. The values of RSDs (n=7) were less than 0. 5%. The recoveries of added standard were 94. 0%-105. 0%.